Catalog Number MC1311P
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an roi-c cage broke during the introduction of the anchoring plate.The plate would not deploy once the cage fractured.The cage and plate were replaced and the procedure was completed.There were no reported patient impacts.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported that an roi-c cage broke during the introduction of the anchoring plate.The plate would not deploy once the cage fractured.The cage and plate were replaced and the procedure was completed.There were no reported patient impacts.
 
Additional Manufacturer Narrative
Corrections in d4: udi number.Additional information in d4: expiration date, h4, and h6: component, investigation type, findings, and conclusions.Device evaluation the product was not returned and no photos were provided, so an evaluation is unable to be performed.Dhr review the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left zimvie¿s control.Potential root cause a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to excess force applied during insertion or incorrect anchor plate trajectory.Device usage: this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-C IMPLANT 14X14 H5
Common Device Name
ROI-C TITANIUM-COATED IMPLANT SYSTEM
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15565067
Report Number3004788213-2022-00075
Device Sequence Number6157953
Product Code OVE
UDI-Device Identifier03662663016063
UDI-Public(01)03662663016063(17)270301(10)C90607
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151934
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMC1311P
Device Lot NumberC90607
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer03/13/2023
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-