Model Number 7590
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.After stent implantation, a 2.50mm x 20mm maverick balloon catheter was advanced for post dilation.However, during inflation at 17 atmospheres for 16-18 seconds, the balloon ruptured.There were no balloon pieces left in the patient's body and completed the procedure with another of same device.There were no patient complications reported and the patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the severely tortuous and severely calcified left anterior descending artery.After stent implantation, a 2.50mm x 20mm maverick balloon catheter was advanced for post dilation.However, during inflation at 17 atmospheres for 16-18 seconds, the balloon ruptured.There were no balloon pieces left in the patient's body and completed the procedure with another of same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a maverick balloon catheter.The hypotube, shaft, tip, and balloon were microscopically and visually examined.There was blood in the inflation lumen, guidewire lumen, and balloon.The balloon was loosely folded.Inspection of the device revealed that there was a 9mm longitudinal tear from the middle to the distal end of the balloon.Product analysis confirmed the reported event, as the balloon had a longitudinal tear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15565430
Report Number2124215-2022-40249
Device Sequence Number2796621
Product Code LOX
UDI-Device Identifier08714729370185
UDI-Public08714729370185
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model Number7590
Device Catalogue Number7590
Device Lot Number0027412229
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer12/06/2022
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-