• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. POLYAXIAL SCREW (DIA. 3.5 MM X 28 MM L); VIRAGE OCT SPINAL FIXATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. POLYAXIAL SCREW (DIA. 3.5 MM X 28 MM L); VIRAGE OCT SPINAL FIXATION SYSTEM Back to Search Results
Catalog Number 07.01702.021
Medical Device Problem Codes Fracture (1260); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2022-00220.
 
Event or Problem Description
It was reported that two virage screws broke at the weld and three others cracked at the weld intra-operatively.On one screw, the surgeon pushed on it from above after the closure top was tightened and the lower housing broke into two pieces at the welds.Then the surgeon noticed the other four screws that either broke or cracked.All of the screws were removed and replaced to complete the procedure.There were no reported patient impacts.This is report five of five for this event.
 
Event or Problem Description
It was reported that two virage screws broke at the weld and three others cracked at the weld intra-operatively.On one screw, the surgeon pushed on it from above after the closure top was tightened and the lower housing broke into two pieces at the welds.Then the surgeon noticed the other four screws that either broke or cracked.All of the screws were removed and replaced to complete the procedure.There were no reported patient impacts.This is report five of five for this event.
 
Additional Manufacturer Narrative
Device evaluation: the piece from lot aas was confirmed to have both welds broken, causing the lower housing to break into its two pieces and allowing the screw components to disassemble.The weld penetration depth was confirmed to be greater than 0.010." root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to excess force applied while torquing the set screws.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYAXIAL SCREW (DIA. 3.5 MM X 28 MM L)
Common Device Name
VIRAGE OCT SPINAL FIXATION SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15565656
Report Number3012447612-2022-00224
Device Sequence Number11509163
Product Code KWP
UDI-Device Identifier00889024327146
UDI-Public(01)00889024327146(10)AAS
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133556
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07.01702.021
Device Lot NumberAAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer02/09/2023
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date02/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-