Model Number LF1923 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Unintended Radiation Exposure (4565)
|
Event Date 08/08/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a procedure, the device had a component problem and was damaged while being used on the patient.The insulating material peeled off but not dissolved.Used by connecting to ft10 (serial number unknown).The nurse discovered that the tip was missing during use.It was unknown whether it was from the time of opening or during procedure.The event was reported to the surgeon, and the patient's body and the surgical field were checked, but it could not be found.An x-ray was also confirmed, but it could not be detected in the x-ray, perhaps due to plastic.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had a damaged/missing insulation.It was reported that the device had a component that disengaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: carefully insert and withdraw the instrument through the cannula to avoid damage to the device and/or injury to the patient.Close jaws using device lever before insertion/extraction in the trocar.Use the appropriately sized trocar to allow for easy insertion and extraction of the instrument.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|