PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ZENITION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number ZENITION 50
Medical Device Problem Code Radiation Output Failure (4027)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
It has been reported to philips that the wireless footswitch has not been working since installation.A system restart was performed to reset the base station, however the connection has not been re-established.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Additional Manufacturer Narrative
Section e: initial reporter corrected.
 
Additional Manufacturer Narrative
Philips has investigated this complaint.As per the information collected, the wireless footswitch had never been used.It did not work since installation approximately 6 months prior.The system was not in clinical use when the issue was identified.Philips has confirmed that the reported failure is due to a connectivity issue.Due to a firmware bug, the wireless foot switch can suddenly stop responding when a number of ambient conditions coexist, such as emc disturbance and the presence of other wireless devices in the room.Philips has initiated a medical device correction (2021-igt-pun-011).The customer has a wired footswitch available for use.Codes were updated based on the investigation outcome.
 
Additional Manufacturer Narrative
The reported event occurred prior to completion of the field correction 3003768277-2021/11/22-010-c, which addresses the causes associated with the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITION
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15566684
Report Number3003768277-2022-00463
Device Sequence Number6158521
Product Code OWB
UDI-Device Identifier00884838091535
UDI-Public00884838091535
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K183101
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Remedial Action Repair
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZENITION 50
Device Catalogue Number718096
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer09/09/2022
09/09/2022
09/24/2024
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date11/03/2022
11/18/2022
02/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/20/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3003768277-2021/11/22-01
Patient Sequence Number1
Patient SexUnknown
-
-