• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 12MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS EVOLVE 4.0MM X 12MM - PMAS; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number FDS40012
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vasoconstriction (2126)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is implanted inside the patient's vasculature.
 
Event or Problem Description
It was reported in a clinical trial that, the patient was found with incidental vasospasm during procedure for which 100 ug nitroglycerine given and vasospasm promptly resolved and the procedure was completed successfully.The cec (clinical events committee) has assessed the vasospasm to be causally related to the procedure and possibly related to subject flow diverter stent (study device).
 
Additional Manufacturer Narrative
Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Based on the information provided in the complaint, the condition being treated was intracranial aneurysm.The subject flow diverter was implanted and the procedure was completed successfully with no device deficiencies reported.An assignable cause of anticipated procedural complication will be assigned to the as reported 'patient vasospasm serious¿ as a product related root cause does not apply and the issue is due to a known physiological effect of the procedure and/or patient condition noted with the directions for use, product labeling and/or risk documentation files.
 
Event or Problem Description
It was reported in a clinical trial that, the patient was found with incidental vasospasm during procedure for which 100 ug nitroglycerine given and vasospasm promptly resolved and the procedure was completed successfully.The cec (clinical events committee) has assessed the vasospasm to be causally related to the procedure and possibly related to subject flow diverter stent (study device).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 4.0MM X 12MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15566892
Report Number3008881809-2022-00497
Device Sequence Number20352931
Product Code OUT
UDI-Device Identifier07613327386516
UDI-Public07613327386516
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/26/2024
Device Model NumberFDS40012
Device Catalogue NumberFDS40012
Device Lot Number22888004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer11/21/2022
Initial Report FDA Received Date10/07/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AXS CATALYST 5 INTERMEDIATE CATHETER (STRYKER); EXCELSIOR XT-27 MICROCATHETER (STRYKER)
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age39 YR
Patient SexFemale
Patient Weight61 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-