On (b)(6) 2021, a patient participating in a study received treatment for a thoracoabdominal aortic aneurysm.During treatment, a two 7 x 59 gore® viabahn® vbx balloon expandable endoprosthesis (vbx device) were overlapped and implanted in left renal artery.One vbx device was implanted most proximal within the portal and vbx device was implanted as first distal extension.On (b)(6) 2022, patient follow-up surveillance revealed high renal velocities bilaterally and the surveillance ct scan suggested a left renal stenosis distal to the branch.Angiogram revealed no right renal stenosis.On (b)(6) 2022, a reintervention was performed due to the distal left renal artery stenotic lesion.No renal function deterioration was observed.Patient tolerated the procedure.
|
|
Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Review of device manufacturing record history confirmed device met pre-release specifications.Device remains implanted; therefore, direct product analysis was not possible.Instructions for use for gore® viabahn® vbx balloon expandable endoprosthesis hazards and adverse events procedure related: as with all procedures that utilize techniques for introducing a catheter into a vessel, complications may be expected.These complications include, but are not limited to: access site infection; entry site bleeding and/or hematoma; vessel thrombosis, occlusion, pseudoaneurysm, and trauma to the vessel wall (including rupture or dissection); distal embolization; arteriovenous fistula formation; transient or permanent contrast induced renal failure; renal toxicity; sepsis; shock; radiation injury; myocardial infarction; fever; pain; malposition; malpositions; inflammation; and/or death.Device related: complications and adverse events can occur when using any endovascular device.These complications include, but are not limited to: hematoma; stenosis, thrombosis or occlusion; distal embolism; side branch occlusion; vessel wall trauma and/or rupture; false aneurysm; infection; inflammation; fever and/or pain in the absence of infection; deployment failure; migration; and device failure.A possible complication which may occur in conjunction with the use of any heparin-containing product: hit type ii.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|