| Model Number |
SR-0830-CS |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
09/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product associated with this complaint was not returned to the manufacturer for analysis.A review of the manufacturing records for this device was completed and no issues were identified that could have led to the adverse event reported.Complaints will continue to be reviewed and monitored for trends and a supplemental mdr will be submitted if additional information is received.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 24 hours after the completion of a transcarotid artery revascularization (tcar) procedure, the patient experienced a transient ischemic attack (tia).Computed tomography angiography (cta) revealed that the proximal portion of the enroute stent appeared to be subintimal.The physician elected to explant the stent and perform a carotid endarterectomy (cea).The patient was recovering from the cea procedure.No further information was provided.
|
| |
|
Search Alerts/Recalls
|