| Model Number |
VTICM5_13.7 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.7mm vticm5_13.7 -13.50/2.5/098 (sphere/cylinder/axis) implantable collamer lens was missing chunk from periphery and was not implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H3- device evaluation: lens was returned in liquid, in vial.Visual inspection found one of the haptics torn.The torn piece was not returned.H6- device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim # (b)(4).
|
| |
|