| Model Number |
7209161 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during a shoulder arthroscopy, at the time of passing through the ladrum, the tip of the cuff stitch broke, the broken piece got stuck in the instrumental.Surgery was resumed, without any delay, with a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The device is fractured at the distal end of the device.The handle and shaft were returned without the piece that fractured away.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.H11: corrected information in h6 (health effect - impact code).
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The device is fractured at the distal end of the device.The handle and shaft were returned without the piece that fractured away.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states in the absence of the requested supporting medical documentation, clinical factors which could have contributed to the reported adverse event could not be definitively concluded, and a causal relationship between the smith and nephew¿s device and the reported adverse event could not be determined.A single undated, unlabeled image of the device was provided.The cuff stitch are manufactured and intended as externally communicating devices and they are not approved for long term internal tissue exposure and long-term implantation data is not available.If the tip was retained, the patient impact beyond possible micro-motion and/or migration, and local irritation/discomfort cannot be determined.However, we cannot make conclusions on the impact of the non-implantable foreign body.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|