| Catalog Number |
07.02054.001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of a vitality screwdriver was found fractured when putting the kit back together after surgery.There were no impacts during the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection revealed that the tip of the driver with lot # ab79 was fractured and twisted.Functional testing revealed that the driver was unable to mate with a vitality screw.Potential cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.It could also be attributed to off-axis forces applied during use.Dhr review : per dhr review, the parts were likely conforming when it left zimmer biomet control.Device use: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of a vitality screwdriver was found fractured when putting the kit back together after surgery.There were no impacts during the surgery.
|
| |
|