• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 36 +3 / 33 COMBO HUMERAL INSERT; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 36 +3 / 33 COMBO HUMERAL INSERT; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number 36 +3 / 33 COMBO HUMERAL INSERT
Health Effect - Clinical Code Pain (1994)
Date of Event 01/22/2021
Type of Reportable Event Serious Injury
Event or Problem Description
On 08/04/2022, it was reported by a patient via email that she underwent a reverse shoulder replacement on (b)(6) 2021, and is now experiencing sudden pain to her right shoulder when raising her arm overhead.X-rays suggest a stress fatigue fracture through the inferior screw.This demonstrates loosening of the left glenoshpere/baseplate.No additional information provided.Additional information requested.Additional information provided on 01/22/2021: implanted devices are as follows: right reverse tsr, arthrex univers 6mm stem with 36/+2 right suture cup, small 33/+3 humeral insert, 24mm mono-block base plate with 3 locking screws,+4/33mm glenophere.A ct scan of the right shoulder obtained at lehigh valley hospital is available for review.This demonstrates appropriate positioning of reverse tsr components.A minimal amount of fluid is seen.There is degenerative change to the ac joint.No lucency around the implants.Due to the artifact the screw fracture is not visible.At this time a revision surgery will not take place.Additional information requested.Additional information provided (b)(6) 2022: implanted devices are as follows: ar-9560-24p, batch 8164; ar-9502f-36rcpc, batch: 19.04738; ar-9564-2433-lat, batch: 19.00839; ar-9503-3633-3, batch: 19.02941; ar-9563-20, batch: 2019004605; ar-9563-32, batch: 2019004423; ar-9562-32nl, batch: 2019002834; ar-9501-06p, batch: 19.04248.It was reported by the patient that the revision took place on (b)(6) 2022.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36 +3 / 33 COMBO HUMERAL INSERT
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15569651
Report Number1220246-2022-05585
Device Sequence Number2797355
Product Code PHX
UDI-Device Identifier00888867296596
UDI-Public00888867296596
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number36 +3 / 33 COMBO HUMERAL INSERT
Device Catalogue NumberAR-9503-3633-3
Device Lot Number19.02941
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date10/10/2022
Date Device Manufactured01/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-