• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-52
Device Problem Device Displays Incorrect Message (2591)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit failed on start-up error 14.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit failed on start-up error 14.There was no patient involvement reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the unit final check being made after new pcb and helium line fitted for previous call numbers for return to customer.Unit alarmed and error 14 showing on start up.This is a faulty compressor fault so replaced the compressor (d119-00-0236) and retested unit.The equipment was cleared for customer use.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15570435
MDR Text Key306522432
Report Number2249723-2022-02604
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108438
UDI-Public10607567108438
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/30/2022
Initial Date FDA Received10/10/2022
Supplement Dates Manufacturer Received04/12/2023
02/22/2024
Supplement Dates FDA Received04/24/2023
02/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-