• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number A1030-060
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty advancing and removing the device from the guide wire.Factors that may contribute to the difficult advancing and removing the balloon dilatation catheter (bdc) from the guide wire causing resistance between the devices may include, but not limited to, manufacturing damage, device placement technique, guiding catheter support, inner diameter of guide wire lumen, condition of the guide wire, or damage to the distal shaft of the catheter.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.The additional ht command device referenced in b5 is filed under separate medwatch report number.Na.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified tibial artery that is 90% stenosed.The 014 ht command guide wire crossed lesion successfully.The 3.0x60mm armada 14 balloon dilatation catheter was advanced and used to pre-dilate the lesion successfully even though resistance was met during advancement with the guide wire.Once removed the 4.0x38mm xience prime stent delivery system was attempted to advance over the wire; however, resistance was met with the guide wire.The same armada 14 balloon was attempted to be advanced over the guide wire, but met resistance was noted again.The balloon and guide wire were removed together.The guide wire was removed and the coating was noted to be peeled but remaining on the guide wire.Another command guide wire and xience prime were used to successfully complete the procedure.There were no clinically significant delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15571248
Report Number2024168-2022-10458
Device Sequence Number2906418
Product Code DQY
UDI-Device Identifier08717648155802
UDI-Public08717648155802
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberA1030-060
Device Catalogue NumberA1030-060
Device Lot Number1083141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/15/2022
Initial Report FDA Received Date10/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
014 HT COMMAND GUIDE WIRE
Patient SexUnknown
-
-