• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NISSHA MEDICAL TECHNOLOGIES LTD BOSTON SCIENTIFIC; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NISSHA MEDICAL TECHNOLOGIES LTD BOSTON SCIENTIFIC; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number DGP-PM2-25
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 09/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Approximated based on the date the manufacturer became aware of the event.
 
Event or Problem Description
It was reported that the patient received a burn from the grounding pads during a genicular rf procedure.It was noted that the burn was first degree outer edges and second degree with yellow slough in the center.The patient was brought to emergency room to treat the burn.The patient was discharged and was prescribed silvadene dressing treatments.
 
Additional Manufacturer Narrative
Dgp-pm2-25 ln: 40024.The returned grounding pad was analyzed and passed visual inspection.However, it was not specified whether the area where the grounding pad was placed was shaved and clear of any lotion or cream.Thus, the application site was likely not prepared correctly prior to application.
 
Event or Problem Description
It was reported that the patient received a burn from the grounding pads during a genicular rf procedure.It was noted that the burn was first degree outer edges and second degree with yellow slough in the center.The patient was brought to emergency room to treat the burn.The patient was discharged and was prescribed silvadene dressing treatments.
 
Additional Manufacturer Narrative
This 3500a mdr was submitted in error.The regulatory reporting for this device is the responsibility of the manufacturer (nissha medical technologies ltd.) and not boston scientific.Therefore, please note that this event should not have been reported on a 3500a mdr form by boston scientific.
 
Event or Problem Description
It was reported that the patient received a burn from the grounding pads during a genicular rf procedure.It was noted that the burn was first degree outer edges and second degree with yellow slough in the center.The patient was brought to emergency room to treat the burn.The patient was discharged and was prescribed silvadene dressing treatments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOSTON SCIENTIFIC
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
NISSHA MEDICAL TECHNOLOGIES LTD
torbay business park
woodview road
paignton DE TQ4 7 HP
UK  TQ4 7HP
Manufacturer (Section G)
CEA MEDICAL MANUFACTURING INC
1735 merchants court
colorado springs CO 80916
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key15571352
Report Number3006630150-2022-05345
Device Sequence Number20378571
Product Code GEI
UDI-Device Identifier35060441917622
UDI-Public35060441917622
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/14/2023
Device Model NumberDGP-PM2-25
Device Catalogue NumberDGP-PM2-25
Device Lot Number40024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer11/03/2022
09/12/2024
Initial Report FDA Received Date10/10/2022
Supplement Report FDA Received Date11/24/2022
09/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
-
-