BOSTON SCIENTIFIC CORPORATION WALLFLEX BILIARY; BILIARY STENT SYSTEM FOR BENIGN STRICTURES Back to Search Results
Model Number M00570360
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Liver Damage/Dysfunction (1954)
Date of Event 09/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022, that a wallflex biliary rx fully covered rmv stent was implanted in the common bile duct to treat an approximately 3cm malignant stricture during an endoscopic retrograde cholangiopancreatography (ercp) procedure with stent placement procedure performed on (b)(6) 2022.The patient's anatomy was tight and was not dilated prior to stent placement.On (b)(6) 2022, during an emergent follow-up, the patient presented with elevated bilirubin level (abnormal liver function test and/or blood count).An ercp procedure was performed, and it was found that the wallflex biliary stent had migrated distally out of the common bile duct and was unable to be located.According to the physician, the stricture was not resolved at the time of migration.A plastic stent was implanted for biliary decompression and the procedure was completed.The patient was reported to be fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX BILIARY
Common Device Name
BILIARY STENT SYSTEM FOR BENIGN STRICTURES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15571415
Report Number3005099803-2022-05958
Device Sequence Number11458467
Product Code PNB
UDI-Device Identifier08714729953173
UDI-Public08714729953173
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
DEN150040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00570360
Device Catalogue Number7036
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date10/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
Patient RaceWhite
-
-