Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Report source: foreign: australia.Concomitant medical products: e1 44-36 std hmrl brng cat: ep-115393 lot: 618100.Comp rvs tray co 44mm cat: 115370 lot: 467430.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event or Problem Description
It was reported patient underwent a revision procedure approximately 2 years and 7 months post implantation due to loosening of humeral stem following cancer treatment.No additional patient consequences were reported.Attempts have been made and additional information on the reported event is unavailable.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Proposed code: mechanical (g04) - stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional.Image was not sent for further review due to poor quality and undated image.Device history record was reviewed and no discrepancies were found.It was stated that the loosening occurred after the patient underwent cancer treatments.However, as no further medical record were provided, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP PRIMARY STEM 8MM MINI
Common Device Name
SHOULDER, PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15573160
Report Number0001825034-2022-02286
Device Sequence Number20378444
Product Code HSD
UDI-Device Identifier00880304449756
UDI-Public(01)00880304449756(17)280611(10)771760
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K193038
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number113628
Device Lot Number771760
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date10/10/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/11/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight64 KG
-
-