Model Number PCE60A
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #331a54.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances / manufacturing irregularities were identified.As part of ethicon's quality process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the surgery, after fired, noted some staples malformed with irregular shape.Used suture to oversew.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 11/3/2022.D4: batch #331a54.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pce60a device was returned with no apparent damage and with an ecr60w reload partially loaded on the device.The reload was received partially fired 1/16.The returned device and reload were tested for functionality in the straight position by resetting and reloading it into the device.The device achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment slot resulting in the knife pushing the one-piece sled forward and locking the reload.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PWRD ECH FLEX 60MM 28CM SHAFT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15573393
Report Number3005075853-2022-06771
Device Sequence Number6077291
Product Code GDW
UDI-Device Identifier10705036014478
UDI-Public10705036014478
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberPCE60A
Device Catalogue NumberPCE60A
Device Lot NumberV9553F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/16/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date10/10/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ECR60W
Patient SexUnknown
-
-