| Model Number |
39540 |
| Medical Device Problem Codes |
Material Perforation (2205); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a hole in the catheter occurred.The totally occluded target lesion was located in the non-tortuous and calcified superficial femoral artery.A 135cm rubicon 35 catheter was selected for use.During an attempt to get through the occlusion, it was noted that the distal end of the catheter folded onto itself, and the wire poked a hole through the catheter.The device was pulled and removed intact from the patients body.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the complaint device was received for product analysis.The device was visually and microscopically inspected for damage.The devices shaft showed damage in the form of buckling/kinks located 135cm from the hub.There was a hole in the catheter shaft located 134.5cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that a hole in the catheter occurred.The totally occluded target lesion was located in the non-tortuous and calcified superficial femoral artery.A 135cm rubicon 35 catheter was selected for use.During an attempt to get through the occlusion, it was noted that the distal end of the catheter folded onto itself, and the wire poked a hole through the catheter.The device was pulled and removed intact from the patients body.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|