Model Number 39540
Medical Device Problem Codes Material Perforation (2205); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a hole in the catheter occurred.The totally occluded target lesion was located in the non-tortuous and calcified superficial femoral artery.A 135cm rubicon 35 catheter was selected for use.During an attempt to get through the occlusion, it was noted that the distal end of the catheter folded onto itself, and the wire poked a hole through the catheter.The device was pulled and removed intact from the patients body.The procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.The device was visually and microscopically inspected for damage.The devices shaft showed damage in the form of buckling/kinks located 135cm from the hub.There was a hole in the catheter shaft located 134.5cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that a hole in the catheter occurred.The totally occluded target lesion was located in the non-tortuous and calcified superficial femoral artery.A 135cm rubicon 35 catheter was selected for use.During an attempt to get through the occlusion, it was noted that the distal end of the catheter folded onto itself, and the wire poked a hole through the catheter.The device was pulled and removed intact from the patients body.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBICON 35
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15574359
Report Number2124215-2022-40345
Device Sequence Number2797508
Product Code DQY
UDI-Device Identifier08714729832072
UDI-Public08714729832072
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K122394
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/18/2025
Device Model Number39540
Device Catalogue Number39540
Device Lot Number0029794642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer10/17/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
Patient RaceBlack Or African American
-
-