INFUTRONIX, LLC. ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP; IV ADMINISTRATION SET Back to Search Results
Model Number HS-001
Medical Device Problem Codes Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A distributor of infutronix received a complaint from a healthcare professional (hcp) reporting "the admin set tubing fell apart".Tubing disconnected from the cassette at the proximal end, and leaked.Patient returned to the facility to get further attention and to return the defective tubing.Device operator was the patient.Medication being infused was unknown.No patient injury reported.The contract manufacturer of the affected device is podo xingda (tianjin) medical co.Ltd.
 
Additional Manufacturer Narrative
Infutronix is waiting for the affected device to be returned.Further information from the reporter regarding event, product, and patient details has been requested.No additional information is available at this time.
 
Additional Manufacturer Narrative
Device evaluation: a visual inspection was performed on the hs-001 (lot unknown) administration set, and it was observed that the tubing was separated from the proximal end of the cassette.The report complaint of "admin set tubing fell apart" has been confirmed.The possible cause of the leaking may be a bonding issue on the proximal end of the cassette.Capa initiated: it2021-17.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADMINISTRATION SET FOR HALO II AMBULATORY INFUSION PUMP
Common Device Name
IV ADMINISTRATION SET
Manufacturer (Section D)
INFUTRONIX, LLC.
177 pine street
natick MA 01760
Manufacturer (Section G)
INFUTRONIX, LLC.
177 pine street
natick MA 01760
Manufacturer Contact
lorraine hanley
177 pine street
natick, MA 01760
5083158230
MDR Report Key15574433
Report Number3011581906-2022-00193
Device Sequence Number6077331
Product Code FPA
UDI-Device Identifier00817170020017
UDI-Public00817170020017
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K153193
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberHS-001
Device Catalogue NumberHS-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown