EXACTECH, INC. PS TIBIAL INSERTS SZ 5, 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
PS TIBIAL INSERTS SZ 5, 15MM |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
11/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, approximately two years post initial tka, patient had a revision due to prosthesis wear.No additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) the revision reported may have been the result of prosthesis wear and/or inclusion of the polyethylene component in the packaging recall.Additionally, a contributing factor to the wear may have been result of being packaged in a non-conforming vacuum bag for more than five years.However, the prosthesis wear cannot be confirmed as the devices were not available for evaluation and no additional clinical information was provided.The most probable root cause associated with the reported event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b3, d4 (expiration date), d6b, e4 (remove erroneous entry from initial report), g2 (brazil), h4, h6 (clinical code, component code, investigation findings, investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|