Model Number PS TIBIAL INSERTS SZ 5, 15MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 11/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, approximately two years post initial tka, patient had a revision due to prosthesis wear.No additional information available.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) the revision reported may have been the result of prosthesis wear and/or inclusion of the polyethylene component in the packaging recall.Additionally, a contributing factor to the wear may have been result of being packaged in a non-conforming vacuum bag for more than five years.However, the prosthesis wear cannot be confirmed as the devices were not available for evaluation and no additional clinical information was provided.The most probable root cause associated with the reported event of ¿prosthesis wear¿ is associated with material damage to a surface, usually involving progressive loss or displacement of material, due to relative motion between that surface and a contacting substance or substances.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b3, d4 (expiration date), d6b, e4 (remove erroneous entry from initial report), g2 (brazil), h4, h6 (clinical code, component code, investigation findings, investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 5, 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
ashley harris
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key15574563
Report Number1038671-2022-01273
Device Sequence Number2984587
Product Code JWH
UDI-Device Identifier10885862048493
UDI-Public10885862048493
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/13/2021
Device Model NumberPS TIBIAL INSERTS SZ 5, 15MM
Device Catalogue Number204-25-15
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer01/27/2023
03/09/2026
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date02/16/2023
03/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/24/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-