ZIMMER SWITZERLAND MANUFACTURING GMBH CMN FEMORAL NAIL, CCD 125, LEFT, 10 MM, 21.5 CM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Failure of Implant (1924); Pain (1994)
|
| Date of Event |
09/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a surgery for fixation of a pertrochanteric fracture of the left femur, and approximately 2 months later the implant was found to be fractured in the bone.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Report source - foreign: italy.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient sustained a fall at home resulting in a severely displaced and comminuted left subtrochanteric femoral neck fracture, a reduction with im nail internal fixation was performed without complications.The patient did well and showed signs of fracture consolidation in follow up x-rays.Approximately 5 months postop, the patient presented to the emergency department due to severe pain after discontinuing ambulating with assistance a few days prior.Ct showed multiple subtrochanteric fractures of the left femur with detachment of the lesser trochanter and angulation of the proximal portion of the nail.The patient was revised revision post implantation; during which, the nail was found broken above the cephalic screw.Poor bone quality was noted with absence of osteoreparative callus at the site of the previous fracture with consequent hypermobility of the fracture stumps.The nail and screws were removed, and a competitor total hip arthroplasty was placed.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: a2, a3, b3, b4, b5, b7, d6b, g3, g6, h2, h6, h11.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, d10, g3, g6, h2, h3, h6, h11.D10: item # 47248403050, 5.0 mm diameter cortical screw - red fixed angle 3.5 mm hex head, lot # 65267042.Item # 47248510510, znnâ¿¢, cmn lag screw, 10.5 mm, 105 mm including set screw, lot # 3038603.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Review of the complaint history identified additional similar complaints for the reported item and no additional similar complaints for the reported part and lot combination.Medical records for the implantation were provided and reviewed by a health care professional.The review identified the patient sustaining a closed subtrochanteric fracture of the left femur after an accidental fall at home.Im nail and anti-rotational screw placed successfully with no intraoperative or postoperative complications.Medical records for and leading up to the revision dated were provided and reviewed by a health care professional.The review identified that the patient with poor bone quality initially showed good fracture consolidation and mobility but later developed complications leading to revision surgery.Indication for revision surgery was given due to subtrochanteric fracture and nail breakage, with the nail found being broken above the cephalic screw.Competitor tha placed.Based on the available information, the reported event can be confirmed.The medical records indicate that the patient had poor bone quality, which may have contributed to the reported issue.However, based on the available information and investigation results, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|