Batch review performed on 21 september 2022: lot 2114131: 110 items manufactured and released on 10-dec-2021.Expiration date: 2026-11-23.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event in the period of review.Additional device involved: batch review performed on 21 september 2022: mectacer 01.29.204 biolox delta ceramic ball head 12/14 ø 32 size s -4 (k112115) lot 2111080: 200 items manufactured and released on 06-dec-2021.Expiration date: 2026-11-21.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event in the period of review.Visual inspection performed by medacta r&d hip project manager on (b)(6) 2022: from the received part, it was possible to note a slight abrasion on the bottom of the inner surface od the poly liner.Morevover, a lipidic absorbment was noticed from the yellowed color of the liner.From the received information, the root cause is unknown, but it is subspected a third body presence.However, a cq analysis was required.Update of (b)(6) 2022.From the received controls, values were found in specifications, so the subspected abrasion was not present or, even if present for a third body presence, the liner was still in specifications.From the received information it is not possible to determine with certainty the root cause of the event, but we can exclude a failure of the device.
|