• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012453-20
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was an issue with deflation with the 4.0x20 mm nc trek balloon dilatation catheter.There were no reported adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device location was not provided, and it is unknown if the device is returning for analysis.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported deflation issue.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported deflation problem was unable to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As the reported deflation problem was unable to be confirmed during return analysis it is possible that interactions with the anatomy/other devices and/or the noted device damages (bunched inner and outer member) resulted in compromising the inflation/deflation lumen; thus resulting in the reported deflation problem; however this cannot be confirmed.The investigation determined a conclusive cause for the reported deflation problem cannot be determined.Interaction/manipulation of the device likely resulted in the noted bunched inner and outer member which possibly contributed to the reported deflation problem.The stretched guide wire exit notch appears to have resulted from an interaction with the guide wire.This type of mechanical damage can occur if an attempt is made to pull the stent delivery system in an opposite direction as the guide wire.The noted multiple bends on the hypotube likely occurred due to handling or during packing for return analysis.There is no indication of a product quality issue with respect to manufacture, design or labeling.H6- type of investigation code 4115 (device discarded) was removed and replaced with code 10 (testing of actual/suspected device) investigation findings code 114 operational problem identified was removed and replaced with code 213 (no device problem found).D9 - device available for eval updated from 'no' to 'yes' h10: addtl mfg narrative ¿ updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15576300
Report Number2024168-2022-10459
Device Sequence Number12307084
Product Code LOX
UDI-Device Identifier08717648152085
UDI-Public08717648152085
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model Number1012453-20
Device Catalogue Number1012453-20
Device Lot Number20511G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/22/2022
Supplement Date Received by Manufacturer10/10/2022
11/15/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date10/13/2022
12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-