ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFNT40
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Visual Disturbances (2140)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Three high quality slit lamp images were provided and reviewed.Two images show the anterior surface of an intraocular lens at high magnification, and the third is a retro-illuminated view.The three images show what appear to be deposits on the anterior surface of the intraocular lens, and the retro-illuminated view additionally shows what appears to be a linear strand running across the approximately lower left quadrant on the posterior surface of the intraocular lens.The source of the observations is unable to be determined based on the images.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that following an intraocular lens (iol) implant procedure, intraocular lens had "dots/marks" on the surface and the patient experienced visual discomfort at night.There are two medical device reports associated with this patient.This report is associated with file 1 of 2.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15576615
Report Number1119421-2022-02154
Device Sequence Number14148303
Product Code MFK
UDI-Device Identifier00380652390514
UDI-Public00380652390514
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/04/2026
Device Model NumberTFNT40
Device Catalogue NumberTFNT40.195
Device Lot Number15145675
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Initial Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-