• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I 3I T3¿ WITH DCD¿ NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3¿ WITH DCD¿ NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number BNST3210
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that implant fractured.Patient received 1 stage implant placement procedure and final attachment restoration.After 2 years of functioning , the implant was suddenly fracture in two pieces., so we had to remove the implant.Tooth site # 15.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
One (1) 3i t3 with dcd non-platform switched tapered implant 3.25 x 10mm (bnst3210) was returned for investigation.Visual evaluation of the as returned implant identified fracture at the threads, signs use, and external damage possibly from the removal process.Zimvie is not responsible for any damage caused by the clinician's removal of the device.Additionally, an unknown zest locator was returned.However, no information was given about this product and no malfunction was identified, so it has been added as an associate item.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Pre-existing conditions noted on the per were smoker, inadequate oral hygiene, and low bone density (type iii).The reported device was in tooth location 15 (fdi) and was used for approximately 1 year, 10 months.The customer did provide x-ray or pictures.Appropriate documents were reviewed.Dhr review was completed for the subject lot number 2018040668.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformance's, which could have caused or contributed to the reported event, were noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.December post market trending was reviewed and there were no actionable events or corrective actions for the reported event or product.No actionable items have been triggered that will affect complaint handling on our end for this month.Therefore, based on the available information, device malfunction did occur, and the reported event was confirmed as physical evaluation identified the fractured implant.The following sections have been updated: b4: date of this report d9: device availability g3: date received by manufacturer g6: checked "follow-up" h2: checked follow-up type h3: changed "no" to "yes" h6: entered evaluation codes h10: added manufacturer narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3¿ WITH DCD¿ NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15577289
Report Number0001038806-2022-01537
Device Sequence Number7384788
Product Code DZE
UDI-Device Identifier00844868029090
UDI-Public(01)00844868029090(17)230426(10)2018040668
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
K122300
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 02/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/26/2023
Device Model NumberBNST3210
Device Catalogue NumberBNST3210
Device Lot Number2018040668
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer02/21/2023
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date02/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexMale
-
-