EXACTECH, INC. PS TIBIAL INSERTS SZ 1, 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
204-21-15 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
12/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending investigation.Concomitant medical devices: 200-04-22 plato tibial cemented with flange 1f/2t and 2f/2t exactech 4910293.(b)(4).
|
| |
|
Event or Problem Description
|
|
As reported, the patient had an initial tka on(b)(6) 2017.The patient was revised on (b)(6) 2021 due to poly wear.No other patient information/medical history reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported may have been the result of malalignment between the femoral and tibial components, third body wear, patient-related conditions, or any combination of these possibilities, which led to polyethylene wear of the tibial insert.A contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed as the device was not returned for evaluation and images/radiographs were unable to be obtained.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported may have been the result of malalignment between the femoral and tibial components, third body wear, patient-related conditions, or any combination of these possibilities, which led to polyethylene wear of the tibial insert.A contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.However, this cannot be confirmed as the device was not returned for evaluation and images/radiographs were unable to be obtained.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|