| Catalog Number |
24355 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A hospital alleged multiple sets of 3m¿ ranger¿ high flow warming sets (model 24355) leaked at the bubble trap while using.No patient or staff injury was alleged.No medical interventions were required.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient specifics are unknown.Date of event was reported to be over the last two months.Products have been discarded; therefore, not returned to 3m¿ for analysis.3m¿ is unable to verify the leak, identify exact location and root cause without the sample.3m¿ will continue to monitor.
|
| |
|
Additional Manufacturer Narrative
|
|
3m has investigated and identified a manufacturing issue with the auto-venting bubble trap that introduces the risk of blood or fluid leaks while priming the sets and/or during fluid administration.There is no alleged injury to the patient or caregiver for this reported incident.3m has initiated a field safety corrective action for lots associated with this issue.3m has implemented a change to address the identified manufacturing issue and will continue to monitor.
|
| |
|
Search Alerts/Recalls
|