• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: 3M HEALTH CARE 3M¿ RANGER¿ FLUID WARMING SET, HIGH FLOW; HIGH FLOW DISPOSABLE TUBING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

3M HEALTH CARE 3M¿ RANGER¿ FLUID WARMING SET, HIGH FLOW; HIGH FLOW DISPOSABLE TUBING Back to Search Results
Catalog Number 24355
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A hospital alleged multiple sets of 3m¿ ranger¿ high flow warming sets (model 24355) leaked at the bubble trap while using.No patient or staff injury was alleged.No medical interventions were required.
 
Additional Manufacturer Narrative
Patient specifics are unknown.Date of event was reported to be over the last two months.Products have been discarded; therefore, not returned to 3m¿ for analysis.3m¿ is unable to verify the leak, identify exact location and root cause without the sample.3m¿ will continue to monitor.
 
Additional Manufacturer Narrative
3m has investigated and identified a manufacturing issue with the auto-venting bubble trap that introduces the risk of blood or fluid leaks while priming the sets and/or during fluid administration.There is no alleged injury to the patient or caregiver for this reported incident.3m has initiated a field safety corrective action for lots associated with this issue.3m has implemented a change to address the identified manufacturing issue and will continue to monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3M¿ RANGER¿ FLUID WARMING SET, HIGH FLOW
Common Device Name
HIGH FLOW DISPOSABLE TUBING
Manufacturer (Section D)
3M HEALTH CARE
2510 conway ave
st. paul MN 55144
Manufacturer (Section G)
NYPRO PRECISION ASSEMBLIES
sor juana ines de la cruz
cd. industrial
tijuana, tijuana baja california 22444
MX   22444
Manufacturer Contact
dianne gibbs
3m center, building 275-5w-06
2510 conway ave
st. paul, MN 55144
6517379117
MDR Report Key15577722
Report Number2110898-2022-00087
Device Sequence Number6801166
Product Code BSB
UDI-Device Identifier00707387781446
UDI-Public00707387781446
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K973741
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/27/2025
Device Catalogue Number24355
Device Lot NumberHX9204
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer11/28/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-