• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC ZINGER GUIDE WIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC ZINGER GUIDE WIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number ZNGRMS180HJ
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
An attempt was made to use one zinger guidewire to treat a lesion.The device was inspected with no issues noted.The device was prepped per the ifu with no issues noted.The lesion was not pre-dilated.The device did not pass through a previously deployed stent.Resistance was not encountered when advancing the device and excessive force was not used during delivery.It was reported that theguidewire detached at the tip.It was observed that the left side of the heart had an obstruction so access through the right side of the heart was carried out through the coronary sinus.Two leads were implanted, one in the right ventricular and one in the left ventricular.Upon completion the zinger guidewire was removed without resistance, however it broke at the tip of the device.It was noted that part of the wire remains in the patient.The patient is reported to be alive with no injury.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: force was not applied during advancement or withdrawal of the zinger device.The zinger device was not kinked or re-straightened during use.Two medtronic leads were implanted.The detached portion was not removed from the patient.At the moment it is not planned to remove the detached portion.Initial reporter phone number provided.One image of the device hoop inside sterile sleeve was provided for review.A section of the guidewire was visible, and damage was noted however in the image provided it was unclear if a detachment had occurred.The lot number: g21a04306 and the device details on the label of the sterile sleeve match what was reported in the complaint file.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Product analysis#: (b)(4): fluoroscopic procedural images provided from the account confirm the wire detachment and the fact that the wire remained in the patient and did not appear to be removed.The mechanism of detachment was not evident from the images.Images confirm that the patient underwent previous valve placement and confirm the presence of leads in the vasculature.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZINGER GUIDE WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
Manufacturer (Section G)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15578526
Report Number1220452-2022-00073
Device Sequence Number6707027
Product Code DQX
UDI-Device Identifier00643169452664
UDI-Public00643169452664
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K140695
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/16/2024
Device Model NumberZNGRMS180HJ
Device Catalogue NumberZNGRMS180HJ
Device Lot NumberG21A04306
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer10/11/2022
11/28/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date11/01/2022
12/13/2022
Date Device Manufactured05/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-