| Model Number |
V3.0 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that an ophthalmic operating console exhibited sudden shutdown without displaying any system messages after a cataract surgery, there was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
The company representative confirmed and replicated the reported event.The power supply as subsequently replaced to address the issue.The system was tested and found to meet product specifications.The manufacturing device history record (dhr) was reviewed.The dhr was completed and reviewed by quality assurance (qa) to ensure that the product was manufactured in compliance with the device master record.Based on qa assessment, the product met specifications.The root cause of the reported event is attributed to the nonconforming power supply.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|