MEDICAL COMPONENTS, INC. 2.6F X 50CM DUAL VASCU-PICC Back to Search Results
Model Number MD17012601
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
Post placement procedure, the line cracked internally and had to be removed from the patient.
 
Additional Manufacturer Narrative
We are currently waiting for the device to be returned for evaluation and additional information.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
The 2.6f vascu-picc was returned for evaluation.The lumen length is 23cm.Visual examination of the device revealed a hole in the lumen just beyond the 1cm mark.The lumen appears to burst outward.A functional evaluation indicated that both lumens leaked when flushed.A great deal of blood was noted in both lumens distal to the burst location.The contract manufacturer conducted a review of the manufacture records for the lot number reported.Their investigation revealed the device was manufactured and inspected according to specification with no non-conformances or abnormalities.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have detected any leaks, holes, or weak spots if it had existed at the time of manufacture.Root cause cannot be determined but is most likely not manufacture related.Cause is most likely due to lumen occlusion and flushing against resistance.The instructions for use (ifu) contain the following potential complication: damage/fracture of catheter the ifu contains the following warnings and precautions: *do not flush against resistance.If the lumen will neither aspirate nor flush, and it has been determined that the catheter is occluded with blood, follow institutional declotting procedure.*do not use infusion equipment which can exceed the working pressure of 1.0bar max/750mmhg (14.5 psi).*only use infusion equipment complying with standards, which do not exceed a shut-off pressure of 1.0bar.*bolus injections should be slow and must not exceed the maximum bolus pressure of 1.2bar/900mmhg (17.4 psi).*small syringes will generate excessive pressure and may damage the catheter.The use of 10cc or larger syringes are recommended.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.6F X 50CM DUAL VASCU-PICC
Common Device Name
VASCU-PICC
Manufacturer (Section D)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer (Section G)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer Contact
lynn winkler
1499 delp drive
harleysville, PA 19438
2152564201
MDR Report Key15579550
Report Number2518902-2022-00058
Device Sequence Number15008693
Product Code LJS
UDI-Device Identifier00884908146493
UDI-Public884908146493
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K102966
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/01/2024
Device Model NumberMD17012601
Device Catalogue NumberMD17012601
Device Lot NumberMPZA600
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date12/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-