• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE PULSE SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE PULSE SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 1881500
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An on-site representative was able to confirm the reported event.Pre-case navigational accuracy checks passed following recalibration.After the case, the imaging system and the pulse system were tested together and found to still be accurate.Two (2) concomitant devices, the drill guide and patient reference array, used during the case were also sent back to the manufacturer for evaluation.Testing found both of these devices to meet relevant specifications and to be functioning as intended.Based on the information obtained, the root cause of the reported event cannot be determined.No additional investigation can be completed.Labeling review: ".Warnings, cautions and precautions: read all instructions and understand all warnings and cautions before using the system or accessories.Failure to do so may lead to serious medical consequences.Refer to the instructions for use accompanying other nuvasive products before use with the system to assure proper and safe use of these devices.If system data acquisition seems inaccurate or if the software application does not initiate or malfunctions during use, and recommended steps to restore the system are not successful, abort use of the system."intra-operative warnings:.Assess navigational accuracy repeatedly throughout a procedure when using a surgical navigation system.A) reconfirm accuracy by positioning the navigated instrument tip on an identifiable anatomical landmark and comparing the actual tip location to that displayed by the system.B) if the stereotaxic navigation system does not appear to be accurate despite troubleshooting (e.G.Resetting the system), do not rely on the navigation system.
 
Event or Problem Description
During a case, following instrument registration and an accuracy check using anatomical landmarks using a pointer, k-wires were placed in the patient.A 2d radiograph radiograph was then taken, at which point the surgeon noted that the k-wires that were placed using surgical navigation looked too low relative to the intended location.An additional array registration was conducted to troubleshoot the inaccuracy; however, the system's suggested trajectories were still inaccurate.An alternate imaging system was brought in to complete the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE PULSE SYSTEM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key15579963
Report Number2031966-2022-00200
Device Sequence Number6461230
Product Code OLO
UDI-Device Identifier00887517990792
UDI-Public887517990792
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K210574
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1881500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2022
Initial Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/04/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PULSE DRILL GUIDEPN 5050267; PULSE SINGLE PATIENT REF ARRAYPN 5050401
Patient SexPrefer Not To Disclose
-
-