W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
RLT311415 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
09/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The field sales associate reported the following: on (b)(6) 2015, the patient was implanted with a gore® excluder® aaa endoprostheses to treat an abdominal aortic aneurysm (infrarenal-zone 9).The patient tolerated the procedure.It was reported that images (modality and date is unknown), revealed a proximal type i endoleak.There was aneurysm sac enlargement (amount of enlargement is unknown).Reportedly the patient had very tortuous anatomy, a severely angulated aortic neck and bilateral renal artery stenosis.There was a reintervention on (b)(6) 2022.The physician implanted three gore® excluder® aaa aortic extender endoprostheses in the proximal seal zone from the original procedure.The physician planned to cover the left renal artery, due to stenosis at the proximal connection to aorta and due to the severely angulated aortic neck.The first implanted aortic extender endoprosthesis (cxa360005/24322716) partially covered the left renal artery and was not long enough to cover the angulation.Two additional aortic extender endoprostheses were implanted for coverage of the angulation.The physician attempted to stent the partially covered left renal artery, but the terumo sheath buckled when advancing into the left renal artery.The left renal artery remained partially covered but was patent.The endoleak was resolved.The patient tolerated the reintervention procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), potential device or procedure related adverse events section includes but is not limited to endoleak.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: code c19 ¿ a review of the manufacturing records for the device verified that the lot met all pre-release specifications.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), potential device or procedure related adverse events section includes but is not limited to endoleak.B3: changed the event date from 9/27/2022 to 9/28/2022.
|
| |
|
Search Alerts/Recalls
|
|
|