A system service check of the medrad® stellant ct injection system (sn (b)(4)) was completed by bayer service on september 15, 2022.The system was found to perform to specifications and has been in use daily following the reported occurrence.The customer was unable to provide the specific disposables used in the reported occurrence for evaluation.In addition, lot numbers were not recorded, therefore testing of retained samples is not possible.Additional applications training was offered to the customer and declined at this time.The following excerpts are contained in the medrad® stellant ct injection system operation manual: expel all trapped air from the syringe(s), connectors, tubing, and catheter before connecting the system to the patient.Before connecting the tubing to the patient, visually inspect the syringe and tubing to confirm that all air is purged.To minimize air embolization risks, ensure that one operator is designated the responsibility of filling the syringe(s).Do not change operators during the procedure.If an operator change must occur, ensure that the new operator verifies that the fluid path is purged of air.This information does not constitute an admission that the device, the company, or its employees caused or contributed to a reportable event.
|