MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that , the cs300 intra-aortic balloon pump (iabp) unit had a fiber optic sensor module failure.There was no patient involvement.
 
Manufacturer Narrative
The fse that encountered the issue replaced the batteries.The fiber-optics then started working.The fse completed the pm with a full calibration.The fse tested the unit and found it to be working to manufacturer's specifications.The iabp was returned to the customer and cleared for clinical use.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) unit had a fiber-optic sensor module failure.There was no patient involvement.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15580354
MDR Text Key304714876
Report Number2249723-2022-02613
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 03/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/11/2022
Supplement Dates Manufacturer Received01/09/2023
03/27/2023
Supplement Dates FDA Received01/30/2023
03/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/14/2010
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-