• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 18325
Device Problem Material Twisted/Bent (2981)
Patient Problems Hyperglycemia (1905); Unspecified Infection (1930)
Event Date 09/30/2022
Event Type  malfunction  
Manufacturer Narrative
According to the complainant the device will not be returned for investigation.We are unable to confirm the bent cannula or to determine if it could have contributed to the reported hyperglycemia.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including confirming the blue soft cannula is inspected for any damage.
 
Event Description
It was reported the patient's blood glucose level rose to 300 mg/dl while wearing the pod between 4 and 24 hours.When removed from the infusion site (abdomen), the pod's cannula was found bent.It was also reported that the patient had an infection.As treatment, the patient changed the pod then gave a correction of bolus around 10 units.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key15580636
MDR Text Key306604573
Report Number3004464228-2022-18915
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385082000051
UDI-Public(01)20385082000051(11)211104(17)230504(10)PD1K11042121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 10/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/04/2023
Device Model Number18325
Device Catalogue NumberBLE-P1-525
Device Lot NumberPD1K11042121
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/04/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexFemale
-
-