• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALIGN TECHNOLOGY, INC. INVISALIGN SYSTEM; ALIGNER, SEQUENTIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALIGN TECHNOLOGY, INC. INVISALIGN SYSTEM; ALIGNER, SEQUENTIAL Back to Search Results
Model Number INVISALIGN SYSTEM - LITE
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Tooth Loss (4831)
Date of Event 04/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The current instructions for use (ifu) contains the following: "precautions - existing dental restorations (e.G: crowns, bridges) may become dislodged and require re-cementation or, in some instances, replacement," "a tooth that has been previously traumatized or significantly restored may be aggravated.In rare instances, the useful life of the tooth may be reduced, the tooth may require additional dental treatment such as endodontic and/or additional restorative work, and/or the tooth may be lost," "the health of the bone and gums which support the teeth may be impaired and aggravated." the treating doctor shared that the potential root cause could be treatment plan/movements/staging/velocity.This event is being filed as an mdr as the patient reported tooth loss (serious injury) and the invisalign product was being used.
 
Event or Problem Description
The patient reported loss of tooth #7.It is unknown if the patient required any medical intervention to alleviate te reported symptom.The patient reported being prescribed antibiotics and anti-inflammatories to alleviate the reported symptom.The treatment was discontinued on 04/01/2019, but the condition has not improved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVISALIGN SYSTEM
Common Device Name
ALIGNER, SEQUENTIAL
Manufacturer (Section D)
ALIGN TECHNOLOGY, INC.
2820 orchard parkway
san jose CA 95134
Manufacturer Contact
aaheli poddar
2820 orchard parkway
san jose, CA 95134
4084701340
MDR Report Key15580980
Report Number2953749-2022-02967
Device Sequence Number20429386
Product Code NXC
UDI-Device Identifier00816063020202
UDI-Public(01)00816063020202(10)0042718026(13)181030(91)0795646613N
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K220287
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberINVISALIGN SYSTEM - LITE
Device Catalogue Number8844
Device Lot Number42718026
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/15/2022
Initial Report FDA Received Date10/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/30/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability;
Patient Age58 YR
Patient SexFemale
-
-