• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT SRL ALINIQ AMS; CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT SRL ALINIQ AMS; CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE Back to Search Results
Model Number 03R89-48
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely decreased results from aniq ams software for several newborn patients.The ams software has wrong unit of measure 2500 gram entered for apr testing by the abbott field representative (fsr).The apr unit supposed to be in 2.5 kilogram.The abbott field service representative (fsr) entered wrong unit of measure for apr testing.The apr calculation out of those testing c-reactive protein (crp), 1-acid glycoprotein, haptoglobin.The apr calculation is formula derived from c-reactive protein (crp), 1-acid glycoprotein, haptoglobin.The apr calculation used for newborn infection or sepsis.The incorrect apr results reported out of the laboratory, however that formula changed in software by the abbott field service representative (fsr) with correct the unit of measure as 2.5 kilogram.There were no actual results provided for crp, 1-acid glycoprotein, haptoglobin.No further impact to patient management.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
Technical investigations confirmed that the issue was due to a human error performed by the aliniq ams specialist.The issue reported was due to an incorrect configuration of the aliniq ams rule that calculates the apr score: wrong unit of measure (2500) gram entered for hp test where the weight unit is supposed to be in (2.5) kilogram.Therefore, incorrect apr score was calculated and released.To address this scenario configuration change within the aliniq ams rule was performed at the customer site: hp unit of measure for weight has been properly set to (2.5) kilogram.Customer confirmed that 4 incorrect apr results were reported out of the laboratory and sent to the lis.All incorrect results were repeated and correctly released.No impact on neonates¿ health or safety or incorrect medications have been reported.Ticket trending review did not identify any additional complaints similar to the issue described in the current complaint.Device history record review returned no nonconformances or potential nonconformances related to the issue described in the current complaint.A review of the labeling adequately addresses the customer¿s issue.Based on the investigation, no systemic issue or deficiency of the aliniq ams middleware was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINIQ AMS
Common Device Name
CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE
Manufacturer (Section D)
ABBOTT SRL
via giorgio ribotta 9
rome 00144
IT  00144
Manufacturer (Section G)
ABBOTT SRL
via giorgio ribotta 9
rome 00144
IT   00144
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15581110
Report Number3004032053-2022-00007
Device Sequence Number14998427
Product Code JQP
UDI-Device Identifier00380740189914
UDI-Public00380740189914
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03R89-48
Device Catalogue Number03R89-48
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-