MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 12MM X 45CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 12MM X 45CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036121240
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported during the procedure that when the subject coil was repositioned, it prematurely detached inside the catheter while retracting.The physician replaced both devices and the procedure was completed successfully with no clinical consequences to the patient.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added h3 device evaluated by mfg ¿updated h3 summary attached - updated d4 expiration date - added d9 product available to stryker ¿ updated d9 returned to manufacturer on ¿updated the returned product was confirmed to be the product reported in the complaint.During visual inspection, the subject coil delivery wire was noted to be kinked/bent.The main coil was seen broken/fractured, stretched and removed from the microcatheter.The suture was found damaged.Functional inspection was not performed as visual inspection results proved sufficient.Due to the automated manufacturing execution system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The product was returned, and the reported event was not confirmed during analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.Additional information received stated that the device was inspected prior to use and no issue was noticed.After detachment, no portion of the coil protruded out of the aneurysm or into the parent vessel.The device was returned for analysis and the coil delivery wire was found to be kinked/bent, and the main coil was found to be broken/fractured, stretched and removed from the microcatheter.In the case of this complaint it is most likely that when the device encountered difficulties during the clinical procedure, the damage caused to the delivery wire and main coil was due to handling and manipulation of the device.The as analyzed events of main coil broken/fractured during use, main coil stretched and main coil suture damaged will be assigned with an assignable cause of procedural factors as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.An assignable cause of handling damage will be assigned to the as analyzed event of coil delivery wire kinked/bent as the issue is associated with handling of the device during the clinical procedure, upon removal of the device from the packaging, or preparation of the device prior to use.An assignable cause of not confirmed will be assigned to the as reported event f main coil prematurely dethatched/separated during use as the coil was found to be still attached during device analysis.
 
Event or Problem Description
It was reported during the procedure that when the subject coil was repositioned, it prematurely detached inside the catheter while retracting.The physician replaced both devices and the procedure was completed successfully with no clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 12MM X 45CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15581177
Report Number3008881809-2022-00504
Device Sequence Number5978604
Product Code HCG
UDI-Device Identifier04546540711533
UDI-Public04546540711533
Combination Product (Y/N)N
Initial Reporter CountryRP
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/23/2025
Device Model NumberM0036121240
Device Catalogue NumberM0036121240
Device Lot Number23501180
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/21/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR SL-10 MICROCATHETER (STRYKER); EXCELSIOR SL-10 MICROCATHETER (STRYKER)
Patient Age66 YR
Patient SexFemale
-
-