| Model Number |
20E |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker isolated the issue to the user interface pcb assembly.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report a non-critical issue with their device.Upon evaluation of the customer¿s device, stryker observed that the device displayed a white screen.This is an indication that the device may be inoperable and defibrillation therapy would not be available if needed.The device also logged an event code that is associated with a device lockup.There was no patient use associated with the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report a non-critical issue with their device.Upon evaluation of the customer¿s device, stryker observed that the device displayed a white screen.This is an indication that the device may be inoperable and defibrillation therapy would not be available if needed.The device also logged an event code that is associated with a device lockup.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker further evaluated the removed user interface pcb assembly.It was determined that the memory of integrated circuit chip, designator u2 had become corrupt and caused the reported issue.
|
| |
|