• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported, prior to patient contact for an unspecified procedure, a hair was noticed in the packaging of a flexor ansel guiding sheath.The device was not used for the procedure.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary it was reported, prior to patient contact for an unspecified procedure, a hair was noticed in the packaging of a flexor ansel guiding sheath.The device was not used for the procedure.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.Investigation - evaluation reviews of the complaint history, device history record, instructions for use, and quality control procedures and a visual inspection of the device was conducted during the investigation.The device was returned in sealed packaging for investigation.Physical examination of the returned device showed a hair like fiber in the sealed pouch.A document-based investigation evaluation was performed.A review of the device history record found one relevant non-conformance for ¿foreign matter.¿ no additional complaints were received for this product lot.Although this non-conformance was relevant to the reported failure mode, all non-conforming product was scrapped, there are 100% inspections to capture this non-conformance, adequate inspection activities have been established, and there are no other lot related complaints that have been received from the field.At this time, cook concluded that no non-conforming product from this lot exists in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: how supplied: ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened and undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ based on the available information, cook has concluded that the cause of this failure is a manufacturing and quality control deficiency.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15581323
Report Number1820334-2022-01625
Device Sequence Number17759825
Product Code DYB
UDI-Device Identifier00827002441545
UDI-Public(01)00827002441545(17)250118(10)14478664
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/18/2025
Device Model NumberN/A
Device Catalogue NumberKCFW-6.0-18/38-45-RB-ANL0-HC
Device Lot Number14478664
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer11/07/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date11/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-