| Model Number |
S603 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this pacemaker battery was at less than half a year.Boston scientific technical services (ts) suggested to look back to see if health care professional (hcp) can determine when the first the magnet rate was noted and will be about one year from that time frame that device will reach elective replacement indicator (eri).The device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any even slight changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemakers battery was designed to estimate remaining longevity and trigger corresponding device replacement indicators based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned implantable pacemaker was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any even slight changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemakers battery was designed to estimate remaining longevity and trigger corresponding device replacement indicators based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker battery was at less than half a year.Boston scientific technical services (ts) suggested to look back to see if health care professional (hcp) can determine when the first the magnet rate was noted and will be about one year from that time frame that device will reach elective replacement indicator (eri).Additional information was received indicating that this device was electively explanted due to normal battery depletion (nbd) and a new device was placed.No adverse patient effects were reported.Additional information received indicating that this device was explanted.No additional adverse patient effects were reported.
|
| |
|