It was reported that, during an arthroscopy, when trying to put in the biosure regenesorb screw in tibia, it broke.All the broken pieces were removed using suction and grasper.No significant delay was reported, and the procedure was finished with a smith and nephew back up device in the originally drilled bone hole.No further complications were reported.
|
|
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found listed specification requirements for material composition, moisture content, and storage requirements.An evaluation of the customer provided image showed two fragments of a broken screw on top of packaging.Part number 72204412 and lot number 50982803 can be seen on the packaging.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have definitively contributed to the event.No harm has been alleged.Therefore, no further clinical/medical assessment is warranted.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|