| Model Number |
1008197-20 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the hub of the viatrac balloon catheter separated from the catheter as it was removed from the hoop.Another viatrac was used to successfully complete the procedure.There were no patient involvement and no clinically significant delay in eh procedure reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on the returned device.The reported separation was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported separation.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|