MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES C8XX2, GELPORT 120MM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES C8XX2, GELPORT 120MM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number C8XX2
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit is anticipated to return.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Procedure performed: lap hand assist colon resection.Event description: c8xx2 malfunctioned.Additional information received on 27sep2022 via email from [name], materials manager: "ripped while using".No injuries related to the reported concern.Additional information received on 28sep2022 via email from [name], materials manager.: the alexis sheath tore.The surgeon's hand was cupped appropriately and inserted into the device at the time of tear.The alexis sheath was torn from white to green all the way through.The surgeon removed torn sheath and opened another sheath to complete the case.Additional information received on 29sep2022 via email from [name]: product is available for return.Patient status: no injuries related to the reported concern.Type of intervention: the surgeon removed torn sheath and opened another sheath to complete the case.
 
Event or Problem Description
Procedure performed: lap hand assist colon resection.Event description: c8xx2 malfunctioned.Additional information received on 27sep2022 via email from [name], materials manager: "ripped while using." no injuries related to the reported concern.Additional information received on 28sep2022 via email from [name], materials manager.: the alexis sheath tore.The surgeon's hand was cupped appropriately and inserted into the device at the time of tear.The alexis sheath was torn from white to green all the way through.The surgeon removed torn sheath and opened another sheath to complete the case.Additional information received on 29sep2022 via email from [name]: product is available for return.Patient status: no injuries related to the reported concern.Type of intervention: the surgeon removed torn sheath and opened another sheath to complete the case.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of sheath tear.Based on the condition of the returned unit, it is likely that an object or instrument came in contact with the sheath, resulting in a tear or hole.As the sheath experiences stresses during use, it is possible that the hole or tear propagated through the sheath.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C8XX2, GELPORT 120MM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15583656
Report Number2027111-2022-00770
Device Sequence Number12358759
Product Code GCJ
UDI-Device Identifier20607915123912
UDI-Public(01)20607915123912(17)240331(30)01(10)1450700
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K014047
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberC8XX2
Device Catalogue Number700314608
Device Lot Number1450700
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer09/13/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-