The device manufacturer date is unknown since the lot number was not provided.The subject device was discarded at the user facility.The device history record was not able to be reviewed for the subject device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Factory test unit maj-1555 (lot: 1h2716) was tested and attached to the factory test scope (tjf-q190v, sn: (b)(4) and inserted terumo's syringe to check for leaks, but no leaks were identified.Based on the legal manufacturer's investigation, a definitive root cause could not be identified.However, there are three possible causes: -upon attaching maj-155, the housing broke off.-syringe was not sufficiently inserted.-syringe was not straight when inserted.The maj-1555 is not compatible with the lure lock syringe.Olympus will continue to monitor field performance for this device.
|