| Model Number |
PREVUE C (+16.00 D) |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This initial emdr is being submitted to fda as a reportable event that occurred in the usa.Haptic damage is indicated as a potential malfunction related to the iol, as covered under the warnings section of the product's instructions for use (ifu).Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
|
| |
|
Event or Problem Description
|
|
Damaged haptic after implantation.As they inserted the iol they noticed that both leading and trailing haptics had been amuptated.Iol was bisected to remove.Product replaced with another lens immediately during surgery, without difficulties.
|
| |
|
Event or Problem Description
|
|
Damaged haptic after implantation.As they inserted the iol they noticed that both leading and trailing haptics had been amputated.Iol was bisected to remove.Product replaced with another lens immediately during surgery, without difficulties.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up #1 emdr is being submitted to fda for a reportable event that occurred outside of the usa.The report includes corrected information and additional information not available/included in the initial report.Corrected information: h3 - corrected to yes d6b - corrected explant date additional information: g2 - indicated company representative for investigation g6 - type of report - noted as follow-up #1 h2 - type of follow-up - noted for corrected information and additional information h6 - added codes for manufacturer's investigation: type; findings; and conclusion.The product was not returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.No abnormalities were found in production and inspection records of the product.(serial no.: 8cw10356; model: prevue c).There were not any abnormalities on the loop pull strength test record of the material lot.(sotj-g115-02).This case was reported that the ovd used is "unknown." the exact root cause of the event was not determined.However, based on available information, we believe this event was not caused by our product quality.A review of the most recent complaint trending data indicates that no significant trends have been identified at this time and no capa is required as part of the product evaluation.
|
| |
|