Model Number 1006
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that coating issue occurred.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified superficial femoral artery.A savion flx guidewire was advanced for use.However, when placing the catheter over the wire, the wire coating was peeled.The device was removed intact and another savion guidewire was used to complete the procedure.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the guidewire was returned, and no damages were observed.No issues were observed at the coating of the device.The device was soaked in saline solution in order to activate the coating at distal section.When the coating was activated, no issues were identified in the distal tip coating.
 
Event or Problem Description
It was reported that coating issue occurred.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified superficial femoral artery.A savion flx guidewire was advanced for use.However, when placing the catheter over the wire, the wire coating was peeled.The device was removed intact and another savion guidewire was used to complete the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAVION FLX
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15584825
Report Number2124215-2022-38226
Device Sequence Number11444274
Product Code DQX
UDI-Device Identifier08714729978817
UDI-Public08714729978817
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K180726
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2024
Device Model Number1006
Device Catalogue Number1006
Device Lot Number0029255521
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date10/11/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age88 YR
Patient SexFemale
Patient Weight54 KG
Patient RaceBlack Or African American
-
-