| Model Number |
SA60WF |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Photo provided showing implanted iol(intraocular lens) on a monitor screen.There appears to be a dark line in the center of the iol.Reported complaint cannot be confirmed from the provided photo.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that during implantation of an intraocular lens (iol) fibrous tissue (looking like a scratch on the posterior optic) appeared after the iol implantation.The surgery was completed the same day.No information about lens removal or replacement.Additional information was requested and received stating the patient condition is unknown.The lens is still implanted in the patient's eye.
|
| |
|
Additional Manufacturer Narrative
|
|
The complainant indicates the use of methyl as a viscoelastic, which is not qualified for use with associated model.The manufacturer internal reference number is: (b)(4).
|
| |
|