• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60WF
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Photo provided showing implanted iol(intraocular lens) on a monitor screen.There appears to be a dark line in the center of the iol.Reported complaint cannot be confirmed from the provided photo.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during implantation of an intraocular lens (iol) fibrous tissue (looking like a scratch on the posterior optic) appeared after the iol implantation.The surgery was completed the same day.No information about lens removal or replacement.Additional information was requested and received stating the patient condition is unknown.The lens is still implanted in the patient's eye.
 
Additional Manufacturer Narrative
The complainant indicates the use of methyl as a viscoelastic, which is not qualified for use with associated model.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15585139
Report Number9612169-2022-00524
Device Sequence Number2762226
Product Code HQL
UDI-Device Identifier00380652332682
UDI-Public00380652332682
Combination Product (Y/N)N
Initial Reporter CountryEG
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberSA60WF
Device Catalogue NumberSA60WF.210
Device Lot Number21268286
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/16/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
METHYL; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH IOL DELIVERY SYSTEM, INJECTOR, UNSPECIFIED
Patient SexUnknown
-
-